Description
Description
The C250s Nebulizer with Autosampler Probe (70 cm) is designed to deliver ultra-high sensitivity in both normal and cool plasma modes, making it an excellent choice for semiconductor applications.
- Microconcentric PFA s-type nebulizer ensures precise performance
- Demountable PFA uptake line with zero dead volume connection for optimal washout performance
- Robust design allows for backflushing without the need for cleaning tools
- 70 cm uptake line comprised of approximately 60 cm of flexible uptake line and 10 cm polyamide probe
- Encapsulated polyimide probe with 2.8 mm OD and 10.2 cm length, compatible with Agilent I-AS autosampler only
- 6 mm OD body with 4 mm gas port for efficient gas flow
- Free aspirates at 250 µl/min +/-10 µl/min or can be pumped for flexibility
- Can be used with supplied uptake line or directly connected to line from ESI FAST systems
- Gas line connection kit included
This nebulizer supersedes Part #800-1-020-02-00.
Note: Designed for use only with standard sample introduction systems (ambient temp or chilled spray chamber). If a desolvator is used, select a nebulizer type “Nebulizer for Desolvators.”
Specifications:
- UPTAKE RATE: 250 µl/min
- AUTOSAMPLER COMPATIBILITY: AGILENT I-AS
- PROBE MATERIAL: POLYIMIDE
- FOR DESOLVATOR USE: NO
- A Inches (nn): 3.53″ (89.8)
- B Inches (nn): 0.24″ (6.0)
- C Inches (nn): 5.67″ (144.0)
- D Inches (nn): 0.10″ (2.8)
- TUBING LENGTH Inches (nn): 27.56″ (700.0)