Description
Clarkson BOA200-220V
Laboratory Oil Bath 37L 220V
Yamato Scientific BOA200-220V / Daigger EF29443 / Fisher 50-189-8178 / VWR 75876-012 / Thomas 1171T00
|
Description Operation and functions
Safety features
|
| Specifications | ||
| Models | BOA200-115V | BOA310 |
| BOA200-220V | ||
| Temp. Control Range | RT+10~200°C | RT+10~270°C |
| Temp. Adjustment Accuracy | ±0.1°C | |
| Temp. Distribution Accuracy | ±0.2°C | ±0.3°C |
| Max. Temp. Reaching Time | 120 min. | 70 min. |
| Temp. Control | PID control | |
| Sensor | W sensor: Pt100Ω (Temp. adjuster) + K Thermocouple (Overheat prevention function) | |
| Operation Function | Fixed Operation | |
| Stirring Method | Water Jet Stirring | |
| Heater | SUS316 Pipe Heater 2kW | SUS316 Pipe Heater 4.5kW |
| Safety Device | Electric leakage breaker for over current protection, Self diagnostic function, Temp. sensor error, Heater error, Automatic overheat prevention function, Overheat preventor, Independent overheat preventor (Fixed 230°C for BOA200, Fixed 300°C for BOA310), Float switch detector for overflow oil, Float switch detector, Emergency stop switch | |
| Other Functions | Drain valve, Large operation lamp, External alarm output terminal, Temp. output terminal (1~5V, 4~20mA adjustable), RS485 communication function, Key lock function, Power Failure Protection Function, Calibration Offset Function, External Communication Adapter (optional) | |
| Oil Bath Dimensions WxDxH | 296 x 340 x 270 mm | |
| Bath Capacity | 37L | |
| External Dimensions WxDxH | 531 x 520 x 578 mm (oil bath depth 397 mm) | |
| Power Source | AC115V 18.5A AC220V 10A with external transformer |
AC220V 21A with external transformer |
| Weight | ~37kg | |
| Included Accessories | Shelf 1pc., lid 1pc. |
| Catalog Number | Description | Voltage |
| BOA200-115V | Laboratory Oil Bath 37L | 115V |
| BOA200-220V * | Laboratory Oil Bath 37L | 220V |
| BOA310 | Laboratory Oil Bath 37L High Temp | 220V |